Journal of Peking University(Health Sciences) ›› 2018, Vol. 50 ›› Issue (2): 358-363. doi: 10.3969/j.issn.1671-167X.2018.02.026

• Article • Previous Articles     Next Articles

Verification of skin paste electrodes used in wireless polysomnography

MA Yun-dong1, HUANG Dong2, CHEN Yu-feng2, JIANG Hao-yun2, LIU Jun-hua2, SUN Hong-qiang1△, LI Zhi-hong2△   

  1. [1. Peking University Sixth Hospital, Peking University Institute of Mental Health, Key Laboratory of Mental Health, Ministry of Health (Peking University), National Clinical Research Center for Mental Disorders (Peking University Sixth Hospital), Beijing 100191, China; 2. Institute of Microelectronics, National Key Laboratory of Science and Technology on Micro/Nano Fabrication, Peking University, Beijing 100871, China]
  • Online:2018-04-18 Published:2018-04-18
  • Contact: SUN Hong-qiang, LI Zhi-hong E-mail: sunhq@bjmu.edu.cn, zhhli@pku.edu.cn
  • Supported by:
    Supported by the Peking University Seed Fund for MedicineInformation Interdisciplinary Research Project (BMU20160582), the Capital Health Research and Development of Special (2018-1-4111) and the Capital Foundation for Clinical Characteristics and Application Research(Z151100004015054)

Abstract: Objective: To explore an electrode suitable for wireless portable sleep monitoring equipment and analyze the result of the signals of electrooculogram (EOG) and electroencephalography (EEG) collected by this kind of flexible electrodes. Methods: The flexible electrodes were prepared by microelectromechanical systems (MEMS) technology. This kind of electrodes consisted parylene, chromium, and gold. Parylene, the flexible substrate of this kind of flexible electrodes, was of biocompatibility. Between parylene and gold there was an adhesion layer of chromium, which connected parylene and gold tightly. Then the flexible electrodes were stuck to medical adhesive tape. The electrodes were designed and made into a grid to make sure that the medical adhesive tape could tape on the skin tightly, so that the contact impedance between the electrodes and the skin would be reduced. Then the alternating current impedance of the electrode were tested by the CHI660E electrochemical workstation after the electrode was achieved. To make sure that this kind of electrodes could be used in EOG monitoring, the electrodes were connected to a wireless signal acquisition suite containing special biological signal acquisition and digital processing chip to gather different sites around the eyes and the electrical signals of different directions of the eye movements, then analyzed the signal-to-noise ratio of the EOG. At the end, the Philips A6 polysomnography was used to compare the noise amplitude of the EEG signals collected by the flexible electrode and the gold cup electrode. Results: The electrodes stuck to the skin tightly, and these electrodes could collect signals that we wanted while the experiment was performed. The alternating current impedance of the flexible electrode was between 4 kΩ and 13 kΩ while with the frequency of alternating current under 100 Hz, most EEG signal frequencies were at this range. The EOG signals collected by the flexible electrodes were in line with the clinical requirements. The noise amplitude of EEG signals collected by the flexible electrodes was lower than that of the electrical signals collected by the gold cup electrodes. Conclusion: The flexible electrode could be taken into consideration as an alternative electrode for monitoring EOG and EEG signals, and the wireless portable sleep monitoring devices are to be further developed in the future.

Key words: Polysomnography, Micro-electromechanical systems, Electrooculogram, Electroencephalography, Electrodes

CLC Number: 

  • R741.044
[1] Yang HUO,Bing ZHOU,Hong-yan HE,Long ZHAO,Xue-li ZHANG,Jing LI,Yu-hua ZUO,Yu ZHENG,Zheng-hong REN,Fang HAN,Jun ZHANG. Comparison and correlation analysis of sleep parameters between watch-type sleep monitor (Actiwatch) and polysomnography [J]. Journal of Peking University (Health Sciences), 2021, 53(5): 942-945.
[2] Xin MIAO,Hong-shi HUANG,Xiao-qing HU,Hui-juan SHI,Shuang REN,Ying-fang AO. Changes of electroencephalography power spectrum during joint position perception test after anterior cruciate ligament rupture [J]. Journal of Peking University (Health Sciences), 2021, 53(5): 871-876.
[3] ZHANG Shao-xing,YAO Zi-ming,LUAN Sheng,WANG Li,XU Ying. Detecting of obstructive sleep apnea hypopnea syndrome using a multi-parameter pressure sensitive sleep monitor [J]. Journal of Peking University (Health Sciences), 2021, 53(3): 608-612.
[4] GONG Pan, YANG Zhi-xian, XUE Jiao, QIAN Ping, YANG Hai-po, LIU Xiao-yan,BIAN Kai-gui. Application of scalp-recorded high-frequency oscillations in epileptic encephalopathy with continuous spike-and-wave during sleep [J]. Journal of Peking University(Health Sciences), 2018, 50(2): 213-220.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
[1] Author. English Title Test[J]. Journal of Peking University(Health Sciences), 2010, 42(1): 1 -10 .
[2] . [J]. Journal of Peking University(Health Sciences), 2009, 41(2): 188 -191 .
[3] . [J]. Journal of Peking University(Health Sciences), 2009, 41(3): 376 -379 .
[4] . [J]. Journal of Peking University(Health Sciences), 2009, 41(4): 459 -462 .
[5] . [J]. Journal of Peking University(Health Sciences), 2010, 42(1): 82 -84 .
[6] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 333 -336 .
[7] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 337 -340 .
[8] . [J]. Journal of Peking University(Health Sciences), 2007, 39(4): 346 -350 .
[9] . [J]. Journal of Peking University(Health Sciences), 2007, 39(4): 351 -354 .
[10] . [J]. Journal of Peking University(Health Sciences), 2007, 39(4): 361 -364 .