Journal of Peking University(Health Sciences) ›› 2016, Vol. 48 ›› Issue (2): 218-223. doi: 10.3969/j.issn.1671-167X.2016.02.007

• Article • Previous Articles     Next Articles

A new kinematics method of determing elbow rotation axis and evaluation of its feasibility

HAN Wei1, SONG Jian2, WANG Guang-zhi2, DING Hui2, LI Guo-shen3, GONG Mao-qi1, JIANG Xie-yuan1, WANG Man-yi1△   

  1. (1. The Fourth Clinical Medical College of Peking University, Department of Orthopaedics and Traumatology, Beijing Jishuitan Hospital, Beijing 100035, China; 2. Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing 100084, China; 3. Peking University Health Science Center, Beijing 100191, China)
  • Online:2016-04-18 Published:2016-04-18
  • Contact: WANG Man-yi E-mail:wangmanyi_jst@126.com

Abstract:

Objective:To study a new positioning method of elbow external fixation rotation axis, and to evaluate its feasibility. Methods: Four normal adult volunteers and six Sawbone elbow models were brought into this experiment. The kinematic data of five elbow flexion were collected respectively by optical positioning system. The rotation axes of the elbow joints were fitted by the least square method. The kinematic data and fitting results were visually displayed. According to the fitting results, the average moving planes and rotation axes were calculated. Thus, the rotation axes of new kinematic methods were obtained. By using standard clinical methods, the entrance and exit points of rotation axes of six Sawbone elbow models were located under X-ray. And The kirschner wires were placed as the representatives of rotation axes using traditional positioning methods. Then, the entrance point deviation, the exit point deviation and the angle deviation of two kinds of located rotation axes were compared. Results: As to the four volunteers, the indicators represented circular degree and coplanarity of elbow flexion movement tra-jectory of each volunteer were both about 1 mm. All the distance deviations of the moving axes to the ave-rage moving rotation axes of the five volunteers were less than 3 mm. All the angle deviations of the mo-ving axes to the average moving rotation axes of the five volunteers were less than 5°. As to the six Sawbone models, the average entrance point deviations, the average exit point deviations and the average angle deviations of two different rotation axes determined by two kinds of located methods were respectively 1.697 2 mm, 1.838 3 mm and 1.321 7°. All the deviations were very small. They were all in an accep-table range of clinical practice. Conclusion: The values that represent circular degree and coplanarity of volunteer's elbow single curvature movement trajectory are very small. The result shows that the elbow single curvature movement can be regarded as the approximate fixed axis movement. The new method can replace the traditional method in accuracy. It can make up the deficiency of the traditional fixed axis method.

Key words: Elbow, External fixation, Rotation center positioning

CLC Number: 

  • R687.4
[1] LIU Xing-hua, JIANG Xie-yuan, GONG Mao-qi, ZHA Ye-jun. Effect of radiotherapy and indomethacin together in the prevention of recurrence of ectopic ossification around the elbow after resection [J]. Journal of Peking University(Health Sciences), 2016, 48(2): 230-233.
[2] ZHA Ye-jun, JIANG Xie-yuan, GONG Mao-qi. Treatment of the old terrible triad of the elbow without operative history [J]. Journal of Peking University(Health Sciences), 2016, 48(2): 224-229.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
[1] Author. English Title Test[J]. Journal of Peking University(Health Sciences), 2010, 42(1): 1 -10 .
[2] . [J]. Journal of Peking University(Health Sciences), 2009, 41(2): 188 -191 .
[3] . [J]. Journal of Peking University(Health Sciences), 2009, 41(3): 376 -379 .
[4] . [J]. Journal of Peking University(Health Sciences), 2009, 41(4): 459 -462 .
[5] . [J]. Journal of Peking University(Health Sciences), 2010, 42(1): 82 -84 .
[6] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 319 -322 .
[7] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 333 -336 .
[8] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 337 -340 .
[9] . [J]. Journal of Peking University(Health Sciences), 2007, 39(3): 225 -328 .
[10] . [J]. Journal of Peking University(Health Sciences), 2007, 39(4): 346 -350 .