A multi-view stereo vision methodology for digital soft-tissue impressions in fixed implant rehabilitation of edentulous patients

  • Yongtao YANG 1, 2 ,
  • Yuwen TIAN 2 ,
  • Shenyao SHAN 2 ,
  • Wenbo LI 2 ,
  • Xiangyi SHANG 2 ,
  • Yizhen WANG 2 ,
  • Shuwei GUO 1 ,
  • Zixiang GAO 1 ,
  • Aonan WEN 1 ,
  • Yijiao ZHAO , 1, 2, * ,
  • Yong WANG , 1, 2, *
Expand
  • 1. Center of Digital Dentistry, Department of Prosthodontics, Peking University School and Hospital of Stomatology & National Center for Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Research Center of Oral Biomaterials and Digital Medical Devices & Beijing Key Laboratory of Digital Stomatology & NHC Key Laboratory of Digital Stomatology, Beijing 100081, China
  • 2. Institute of Medical Technology, Peking University Health Science Center, Beijing 100191, China
ZHAO Yijiao, e-mail,
WANG Yong, e-mail,

Received date: 2025-10-13

  Online published: 2025-12-10

Supported by

the National Natural Science Foundation of China(82271039)

the National Key Research and Development Program of China(2022YFC2405401)

the Beijing Natural Science Foundation(L232100)

the Beijing Natural Science Foundation(L242132)

Copyright

All rights reserved. Unauthorized reproduction is prohibited.

Abstract

Objective: To explore the methodology and feasibility of reconstructing soft tissue morphology for fixed implant rehabilitation in edentulous patients using multi-view stereo vision technology, and to conduct a preliminary evaluation of the method's in vitro accuracy. Methods: A pair of edentulous resin models were designed and printed, with 6 implant analogs placed in the maxilla and 4 in the mandible. The experimental group (n=10) utilized a self-developed photogrammetric quad-camera system and the automated reconstruction software RealityScan 2.0.1. Self-developed scan bodies were attached to the analogs, and the handheld camera system was used to capture images of the models in vitro. The images were imported into the software to reconstruct the 3D models, and the data were exported as ".stl" files. The control group (n=10) used an intraoral scanner. Scan caps were attached to the analogs, and the models were scanned to generate ".stl" data. Reference data were obtained by scanning the maxillary and mandibular resin models once each with a desktop scanner (EX-PRO). All data were imported into Geomagic Wrap 2021. The root mean square (RMS) was calculated by comparing the 3D morphology of the experimental and control group data against the reference data to represent the magnitude of the 3D morphological deviation and evaluate accuracy. The evaluation was conducted in 4 specific regions: the alveolar ridge, peri-implant soft tissue, buccal, and lingual areas. Results: In the maxilla, the RMS of the experimental group was significantly higher than the control group in the alveolar ridge [(124.89±21.30) μm vs. (53.90±8.93) μm, P < 0.001], peri-implant soft tissue [(157.74±19.13) μm vs. (67.03±3.94) μm, P < 0.001], and lingual areas [(146.01±33.87) μm vs. (46.20±11.19) μm, P < 0.001]. The RMS in the buccal area was lower for the experimental group than the control group [(50.56±8.34) μm vs. (53.83±12.66) μm], but the difference was not statistically significant (P=0.571). In the mandible, the RMS of the experimental group was significantly higher than the control group in the alveolar ridge [(254.04±88.42) μm vs. (58.28±38.96) μm, P < 0.001], peri-implant soft tissue [(165.18±21.30) μm vs. (70.48±28.20) μm, P < 0.001], and lingual areas [(421.75±59.51) μm vs. (54.59±36.77) μm, P < 0.001]. When comparing the buccal and lingual sides, the lingual RMS was significantly higher than the buccal RMS for the experimental group in both the maxilla (P < 0.001) and mandible (P < 0.001). For the control group, the maxillary lingual RMS was significantly lower than the buccal RMS (P < 0.05), while the mandibular lingual RMS was higher than the buccal, but the difference was not statistically significant (P=0.378). Conclusion: The self-developed quad-camera system, combined with multi-view stereo vision reconstruction software, can successfully record the 3D morphology of soft tissue. This study provides a research foundation for the development of extraoral photogrammetric devices capable of simultaneously determining the spatial positions of multiple implant units and acquiring soft tissue morphology.

Cite this article

Yongtao YANG , Yuwen TIAN , Shenyao SHAN , Wenbo LI , Xiangyi SHANG , Yizhen WANG , Shuwei GUO , Zixiang GAO , Aonan WEN , Yijiao ZHAO , Yong WANG . A multi-view stereo vision methodology for digital soft-tissue impressions in fixed implant rehabilitation of edentulous patients[J]. Journal of Peking University(Health Sciences), 2026 , 58(1) : 126 -132 . DOI: 10.19723/j.issn.1671-167X.2026.01.016

无牙颌种植固定修复是牙列缺失患者的重要治疗方式之一,具有良好的固位和稳定作用[1]。精确获取无牙颌种植固定修复印模是口腔临床诊疗中的重要步骤,不仅需要精准定位口腔内多单位种植体的相对空间位置关系,还需准确复制种植体周围和义齿覆盖的软组织形态,为义齿的设计和制作提供参考[2]。印模精度下降可能导致义齿不能就位或被动就位不足,可引起机械并发症,常见的包括义齿或悬臂梁折断、螺丝松动、断裂等,还可能增加菌斑堆积风险[3]。义齿组织面与软组织间隙过小可能造成被动就位不足、组织压痛,间隙过大可能增加清洁难度和美学风险[1]
随着口腔数字化技术的不断发展,基于摄影测量技术的多单位种植体精确测量方法已在无牙颌种植领域广泛应用,可在口外获取种植体的位置信息,常见的设备有ICam4D (IcamRef,Imetric4D Imaging Sàrl,瑞士)、PIC(PIC dental,西班牙)、Oxo Fit(Oxo,西班牙)等[4]。口外摄影测量设备通过大视场记录并解算编码扫描杆上的特征点位置,计算出种植体的位置,但无法直接获取软组织形态,需要使用口内扫描设备获取软组织形态数据[5]。目前,已有口内摄影测量设备可将种植体定位与软组织复制集成于一体,降低了获取完整无牙颌种植固定修复数字化印模成本,但仍需要分步骤扫描[6]
多视图立体视觉(multi-view stereo,MVS)算法能够实现从多视图视角中恢复场景的三维几何形状,完成二维平面到三维立体的转化,是计算机视觉领域的重要研究方向,与摄影测量算法在几何原理和方法上存在交叉融合[7],且两者均可使用图像传感器采集的图像作为原始数据[8],这为同步实现无牙颌多单位种植体定位和软组织记录提供了理论可行性,但目前该方面研究较少。本研究旨在利用自主研发(简称自研)的摄影测量四目相机组和自动化多视图重建软件,初步探索基于MVS的无牙颌种植固定修复软组织重建的方法及其可行性,为自研摄影测量系统能集成多单位种植体定位和软组织记录功能提供基础。

1 材料与方法

1.1 设备及软件

本研究使用的设备和软件包括:本课题组自研口外摄影测量系统及配套扫描杆,摄影测量系统相机组由1个光源和4个160万像素的电荷耦合器件图像传感器组成;口内扫描仪(TRIOS 3,3Shape公司,丹麦),用于扫描种植树脂模型;高精度牙颌模型三维扫描仪DS-EX-PRO(先临三维科技股份有限公司,杭州),用于扫描种植树脂模型;三维树脂打印机(DLP睿逸DLP1080EA,大族激光科技产业集团股份有限公司,深圳)及牙模树脂/带型材料T-MRD-521,用于打印无牙颌种植树脂模型;Nobel ACT复合基台替代体(Nobel Biocare,瑞士),置于树脂模型中模拟种植体;扫描帽(IcamRef,Imetric4D Imaging Sàrl,瑞士),配合牙颌模型三维扫描仪和口内扫描仪使用;细尖爪数显卡尺(111N201,三量精密量仪有限公司,东莞),精度为0.01 mm,用于测量替代体物理参数;聚醚印模材料(Impregum Soft,3M公司,美国),用于模拟人工牙龈形态;自研系统图像采集上位机,用于控制相机组同步拍摄图像;自动化摄影测量软件RealityScan 2.0.1(Epic Games,美国),将相机组拍摄图像重建为三维网格模型,运行于便携式计算机,GPU为GeForce RTX 4060(8G,NVIDIA,美国);正向设计软件Magics 21.0(Mate-rialise,比利时),用于设计虚拟替代体“. stl”数据;逆向设计软件Geomagic Wrap 2021(3D Systems,美国),用于设计虚拟种植模型,处理获得的“. stl”数据;大族激光固体标记控制软件V5.0(大族激光科技产业集团股份有限公司,深圳),用于树脂模型打印前“. stl”数据排版、分层及输出三维打印机可读取的数据类型。

1.2 种植树脂模型的制作

使用北京大学口腔医院技工室提供的标准化阴模灌制无牙颌石膏模型1对,采用牙颌模型三维扫描仪获取石膏模型“. stl”数据。使用细尖爪数显卡尺测量Nobel ACT复合基台替代体,得到上段圆柱直径4.80 mm、高度4.80 mm,下段圆柱直径3.50 mm、高度8.70 mm,于Magics 21.0软件中设计含两段圆柱的数字化替代体。将数字化替代体“. stl”数据导入Geomagic Wrap 2021软件,分别置于上颌石膏模型扫描数据的16、14、12、22、24、26牙位和下颌石膏模型扫描数据的34、32、42、44牙位,使用“布尔(Boolean)功能”,完成一对数字化上、下颌种植模型的设计,打印制作上、下颌种植模型各1个。将替代体装配于模型并稳定固位,使用聚醚印模模拟人工牙龈填充于替代体周围(图 12)。
图1 上颌无牙颌种植模型

Figure 1 Edentulous maxillary implant model

图2 下颌无牙颌种植模型

Figure 2 Edentulous mandibular implant model

1.3 参考数据的采集

将扫描帽安装于替代体上,使用牙颌模型扫描仪采集树脂模型表面形态数据,上、下颌模型各输出1个“.stl”数据,作为形态偏差分析的参考数据。

1.4 对照组数据的采集

手持口内扫描仪采集安装有扫描帽的树脂模型表面形态数据,上、下颌各10次,输出“.stl”数据。扫描路径为右后颊侧—左后颊侧—左后嵴顶—右后嵴顶—右后舌侧—左后舌侧的连续扫描,每次扫描间隔1 min。

1.5 实验组图像采集和三维数据重建

由同一名操作者将自研扫描杆安装于替代体,树脂模型平放于桌面,手持自研相机组从右侧后牙区颊侧缓慢移动至左侧后牙区颊侧(图 3),模拟定位临床患者多单位种植体的扫描路径,通过上位机在此过程中间隔采集12次无牙颌种植模型的原始图像作为1组,每组共4×12张图像,格式为“.bmp”,上、下颌各采集10组;将每组原始图像导入RealityScan 2.0.1软件中,依次使用“对齐图像”“重建模型(高精度)”功能,重建三维网格模型,导出“.stl”数据。
图3 手持自主研发系统相机组扫描示意图

Figure 3 Handheld scanning procedure with the self-developed camera system

1.6 评价指标

计算模型表面三维形态的均方根(root mean square,RMS)作为指标,使用Geomagic Wrap 2021软件计算各组“.stl”数据与参考数据的RMS,以此反映实验组和对照组记录软组织三维形态的正确度。在Geomagic Wrap 2021软件中,钉住参考数据,先后使用“手动注册-n点注册”“全局注册”将实验组和对照组数据配准至参考数据,对参考数据使用“创建边界”功能选择出要进行形态分析的牙槽嵴顶区域,使用“从边界创建”功能创建分析范围曲线,将曲线“投影”至实验组、对照组数据,然后“转为”边界,复制出参考数据,实验组、对照组边界内的数据,最终所有“.stl”数据被裁剪至相同范围,使用“3D比较”功能计算RMS,对于全牙槽嵴顶区域,进一步划分并选择出穿龈区域(蓝色)、颊侧区域(绿色)和舌侧区域(褐色)计算RMS。
RMS值的计算公式为:$ \text{RMS} =\frac{1}{\sqrt{n}} \times \sqrt{\sum\limits_{i=1}^n\left(X_{1, i}-X_{2, i}\right)^2}$,其中,n代表标准扫描帽测量点的数量,X1, i表示一组数据的第i个测量点,X2, i代表测量点X1, i在另一组扫描数据上对应的最近点。
上、下颌各区域边界见图 4图 5。牙槽嵴顶区域远中与最后侧替代体远中边缘距离约7 mm,颊舌侧距离相应替代体边缘均大于4.5 mm,且边界过渡自然连续;种植体周围软组织区域为模拟人工牙龈范围,在牙槽嵴顶区域基础上,使用“抽取曲线”功能辅以手动编辑,确定颊侧区域和舌侧区域分界,曲率敏感度为80。
图4 上颌各区域边界范围

Figure 4 The boundary scope of the respective regions of the maxilla

图5 下颌各区域边界范围

Figure 5 The boundary scope of the respective regions of the mandible

1.7 统计学分析

使用SPSS 26.0软件,采用Shapiro-Wilk检验验证RMS数据是否符合正态分布,采用Levene检验验证RMS数据的方差齐性,对符合正态分布数据以${\bar x}$±s形式表示,非正态分布数据以中位数(四分位距)[mean (interquartile range), M (IQR)]形式表示。若数据符合正态分布且方差齐,则采用配对样本t检验比较实验组和对照组形态正确度的差异;反之,则采用Wilcoxon符号秩检验比较两组的差异。对组内颊舌侧数据,若数据符合正态分布且方差齐,则采用配对样本t检验比较颊舌侧形态正确度的差异;反之,采用Wilcoxon符号秩检验比较颊舌侧差异。检验水准均为双侧α=0.05。

2 结果

上、下颌三维形态偏差分析色谱图见图 6图 7
图6 上颌三维形态偏差分析色谱图

Figure 6 Color map of maxillary 3D morphological deviation analysis

A, alveolar ridge; B, peri-implant soft tissue; C, buccal area; D, lingual area.

图7 下颌三维形态偏差分析色谱图

Figure 7 Color map of mandibular 3D morphological deviation analysis

A, alveolar ridge; B, peri-implant soft tissue; C, buccal area; D, lingual area.

上颌实验结果显示,实验组在牙槽嵴区域、种植体周围软组织区域和舌侧区域的三维形态偏差均显著高于对照组(均P < 0.001);实验组在颊侧区域的三维形态偏差略低于对照组,差异无统计学意义(P=0.571),见表 1
表1 上、下颌实验组与对照组形态正确度的对比

Table 1 Comparison of morphological trueness between experimental group and control group in the maxillary and mandibular

Items Alveolar ridge Peri-implant soft tissue Buccal area Lingual area
Maxillary
  Experimental group/μm, ${\bar x}$±s 124.89±21.30 157.74±19.13 50.56±8.34 146.01±33.87
  Control group/μm, ${\bar x}$±s 53.90±8.93 67.03±3.94 53.83±12.66 46.20±11.19
  t 9.949 15.643 -0.587 8.984
  P < 0.001 < 0.001 0.571 < 0.001
Mandibular
  Experimental group/μm, ${\bar x}$±s 254.04±88.42 165.18±21.30 57.32±8.26 421.75±59.51
  Control group/μm, ${\bar x}$±s 58.28±38.96 70.48±28.20 49.71±49.21 54.59±36.77
  t 7.031 11.170 0.448 17.466
  P < 0.001 < 0.001 0.665 < 0.001
下颌实验结果显示,相比于对照组,实验组在牙槽嵴区域(P < 0.001)、种植体周围软组织区域(P < 0.001)、舌侧区域(P < 0.001)的三维形态偏差均显著增大,颊侧区域的三维形态差异无统计学意义(P=0.665),见表 1
各组模型的颊侧、舌侧区域对比结果见表 2。对比舌侧区域,实验组在上、下颌颊侧区域的三维形态偏差均显著降低(均P < 0.001);在对照组中,上颌舌侧区域三维形态偏差显著低于颊侧区域(P < 0.05),下颌舌侧区域三维形态偏差略低于颊侧区域,差异无统计学意义(P=0.378)。
表2 组内颊侧区域和舌侧区域形态正确度的对比

Table 2 Comparison of morphological trueness between the buccal and lingual areas

Items Experimental group in mandibular Control group in mandibular Experimental group in maxillary Control group in maxillary
Buccal area/μm, ${\bar x}$±s 50.56±8.34 53.83±12.66 57.32±8.26 49.71±49.21
Lingual area/μm, ${\bar x}$±s 146.01±33.87 46.20±11.19 421.75±59.51 54.59±36.77
t -9.476 2.876 -18.262 -0.928
P < 0.001 < 0.05 < 0.001 0.378

3 讨论

本研究采用自研摄影测量相机组配合MVS重建软件,实现了无牙颌种植模型的软组织重建。在临床获取无牙颌种植数字化印模过程中,常采用摄影测量设备获得种植体间空间位置数据,再使用口内扫描仪获得软组织形态数据,将两数据配准后用于数字化义齿设计[9]。在制作无牙颌种植固定修复义齿时,义齿组织面对应的软组织表面形态需要精确地复制为数字化印模,而目前关于软组织表面形态的精度评价研究较少,尚无研究给出符合临床设计使用的正确度阈值。因此,为对比基于二维图像重建无牙颌种植模型三维形态的正确度,本研究选择已被临床广泛接受的口内扫描仪作为对照组,将实验组和对照组分别与参考数据基于全局“. stl”数据进行配准,在所有“. stl”数据中选择出相同的牙槽嵴区域进行三维形态分析,选择范围大于常规种植固定义齿组织面,为更精准体现实验组和对照组在不同区域的正确度差异,将牙槽嵴区域划分为种植体周围软组织区域、颊侧区域和舌侧区域。这种方法重在分析软组织表面形态的重建差异,软组织形态与多单位种植体位置的配准及其精度有待后续研究。已有研究显示[10],以种植体位置作为配准的基准区域,软组织和种植体位置的配准精度受软组织的采集方式影响。
本研究结果显示,相比于牙槽嵴区域、种植体周围软组织区域和舌侧区域,实验组在颊侧区域的三维形态偏差与对照组差异较小,且在上颌模型中低于对照组(P=0.571),在下颌模型中与对照组的差异无统计学意义(P=0.665),这与实验组拍摄路径在颊侧有关。为了模拟口内空间有限、自研相机组在口外拍摄的临床条件,实验组缺少舌侧细节图像,覆盖范围不足,导致不能获取足够的特征点云进行精细重建,特别是在下颌中,实验组舌侧三维形态偏差达到(421.75±59.51) μm,远高于上颌实验组舌侧的(146.01±33.87) μm,其原因是舌侧不仅存在扫描杆的遮蔽作用,且下颌模型牙槽嵴形态相对于上颌模型更加高耸,增加了舌侧被阻挡的范围;而颊侧有细节图像,覆盖范围较大且没有扫描杆的遮蔽,因此重建的三维形态偏差与口内扫描仪的差异降低。种植体周围软组织区域起于种植体基台或扫描杆外缘,止于人工牙龈外缘部分,相比于其余部分,需要更高的正确度以确保义齿就位时与义齿组织面贴合且不产生阻挡,避免增加食物残渣和菌斑堆积的风险,影响义齿被动就位。实验组在上、下颌的种植体周围软组织三维形态偏差均显著大于对照组(P < 0.001),部分原因也在于舌侧被阻挡。色谱图结果显示,前牙区的种植体周围软组织颊舌侧对比更明显,且舌侧的黄色范围(偏差较大区域)也更广泛。此外,上颌对照组的颊侧三维偏差高于舌侧(P < 0.05),这可能与上颌模型舌侧腭皱提供了更多特征有关。
本研究结果显示,通过模拟口外路径拍摄重建的无牙颌种植模型的舌侧正确度不及颊侧。为了解决舌侧被遮挡而产生视野盲区这一问题,有研究提出利用反光镜补偿盲区,以提升点云配准的精度和鲁棒性(robustness),Ke等[11]的研究利用这一方法在体外实现口外大视场扫描(SSLS组)获得无牙颌多单位种植数字化印模,并与口内扫描仪组(IOS组)对比黏膜的三维形态偏差,结果显示SSLS组的形态偏差为33.4(16.9) μm,IOS组的形态偏差为59.8(21.6) μm,两者差异有统计学意义(P < 0.001);该研究中的IOS组形态偏差与本研究相似,而SSLS组的形态偏差低于本研究的实验组,可能是由于该研究对双目相机进行了内、外参数标定,并使用正弦光栅图案投影至物体表面,使其能够更精确地进行点云重建和配准;此外,该研究在进行黏膜形态偏差分析时,将被裁剪至相同范围的数据再次进行了对齐,以往有研究显示,这种方法可能会忽略对齐数据自身的差异,从而使最终三维形态偏差值低于实际值[12-13]
近年来,随着人工智能的快速发展和图像采集设备的更新,利用深度学习网络从二维图像重建三维场景成为计算机视觉领域的热点方向[14]。传统MVS算法可分为基于面片的方法、多边形网格法、体素法和基于深度图的方法,尽管以上方法在场景重建方面的精度尚可,但仍然面临效率低、完整度不高、难以处理弱纹理区域和反射表面等问题,基于深度学习的MVS局部模块改进和整体结构改进可取得更好的应用效果。有研究报道了将神经辐射场(neural radiance fields,NeRF)结合深度学习技术用于无牙颌种植模型表面重建的效果。Li等[15]的研究采用Neuralangelo高保真神经表面重建算法,将智能手机拍摄的无牙颌种植模型环绕视频经过开源程序Colmap预处理后作为输入,经过不同的训练周期后从检查点中提取三维网格数据,与口内扫描仪对比多单位种植体的定位精度,结果显示,经过5 000个训练周期的多单位扫描杆的总体形态偏差均值为67.69 μm,口内扫描仪的形态偏差均值为57.42 μm,两者差异无统计学意义,Li等[15]的研究虽未分析黏膜重建的精度效果,但是为深度学习结合MVS方法在重建无牙颌种植软组织的应用上指出了方向。
目前,有关无牙颌种植软组织印模精度的方法研究多采用体外研究方式,常采用口内扫描仪作为对照组[11, 15],已有研究报道,体外使用口内扫描仪不受体内黏膜、唾液等影响,相较于体内有更高的精度表现,因此将口内扫描仪作为体外研究的对照组具有一定不足,并不能完全反映其在真实体内的应用效果[16]
本研究仍存在一些局限性,体外实验未能将口内实际环境、软组织动度等因素纳入考量,且本研究利用模型和人工牙龈模拟软组织仅能反映附着龈的重建效果,对活动度更大的游离龈和松软牙槽嵴的应用有限。体外实验的模型与患者实际种植情况、颌骨及软组织条件也存在差异,临床上在患者体内应用条件下的重建正确度有待后续开展进一步研究。此外,本研究是探索基于摄影测量设备进行无牙颌种植软组织重建的初步研究,针对舌侧扫描盲区、扫描杆遮挡、口内弱纹理等问题造成的结果不及对照组这一现象,可采用增加反光板、优化相机组和扫描杆尺寸等方式,增加口内图像的拍摄视角,以获得更大的重建范围,还可引入深度学习方法,提升算法在图像弱纹理区域探测和匹配特征点的性能。后续还需要开展多单位种植体定位与软组织记录同步实现的相关研究,以期提高临床诊疗效率,降低使用成本,促进数字化技术在口腔无牙颌种植领域的广泛应用。
综上所述,本研究初步探索实现自研摄影测量相机组配合MVS重建软件记录软组织三维形态,为无牙颌种植口外摄影测量设备同步定位多单位种植体空间位置和获取软组织形态提供了研究基础,可使数据的采集时间更短于口内扫描仪,有望提高临床诊疗效率,降低使用成本,促进数字化技术在口腔无牙颌种植领域的广泛应用。

利益冲突  所有作者均声明不存在利益冲突。

作者贡献声明  杨咏涛:设计研究方案,数据获取和处理,统计学分析,论文撰写;田淯文:协助设计研究方案,数据获取和处理;单珅瑶:提供数据分析思路,统计学分析;李文博:协助数据获取和处理;商相宜:数据处理和统计学分析;王艺蓁:数据整理和分析;郭殊玮:协助数据处理和整理;高梓翔:协助设计研究方案,统计学分析;温奥楠:参与论文撰写和修改;赵一姣:口腔数字化理论指导,提供研究思路与论文修改;王勇:口腔数字化理论指导,总体研究方案制定与论文修改。所有作者均参与论文修改,并对最终文稿进行审读和确认。

1
周永胜. 口腔修复学[M]. 3版 北京: 北京大学医学出版社, 2020.

2
Rutkūnas V, Gedrimiene A, Mischitz I, et al. EPA consensus project paper: Accuracy of photogrammetry devices, intraoral scanners, and conventional techniques for the full-arch implant impressions: A systematic review [J/OL]. Eur J Prosthodont Restor Dent, 2023 [2023-6-13]. https://pubmed.ncbi.nlm.nih.gov/37314199/.

3
Abdelrehim A , Etajuri EA , Sulaiman E , et al. Magnitude of misfit threshold in implant-supported restorations: A systematic review[J]. J Prosthet Dent, 2024, 132 (3): 528- 535.

DOI

4
Hussein MO . Photogrammetry technology in implant dentistry: A systematic review[J]. J Prosthet Dent, 2023, 130 (3): 318- 326.

DOI

5
杨咏涛, 温奥楠, 商相宜, 等. 基于摄影测量技术的无牙颌种植口外扫描系统的研发和精度评价[J]. 中华口腔医学杂志, 2025, 60 (8): 863- 870.

6
Brakoč J , Todorović A , Mangano FG , et al. Accuracy of intraoral photogrammetry versus direct digital implant impressions in the fully edentulous lower jaw: An in vitro study[J]. J Dent, 2025, 156, 105654.

DOI

7
张力, 刘玉轩, 孙洋杰, 等. 数字航空摄影三维重建理论与技术发展综述[J]. 测绘学报, 2022, 51 (7): 1437- 1457.

8
龚健雅, 季顺平. 从摄影测量到计算机视觉[J]. 武汉大学学报(信息科学版), 2017, 42 (11): 1518-1522, 1615.

9
Revilla-León M, Cascos R, Barmak AB, et al. Registration accuracy of soft tissue information scan captured using an intraoral scanner and implant position scan recorded using extraoral and intraoral photogrammetry systems [J/OL]. J Prosthet Dent, 2025: S0022-S3913(25)00392-0 [2025-05-30]. https://pubmed.ncbi.nlm.nih.gov/40450447/.

10
Revilla-León M , Cascos R , Barmak AB , et al. Alignment accuracy of soft tissue and implant position scans captured by using four intraoral scanners and an intraoral photogrammetry system with and without coded healing abutments[J]. J Prosthet Dent, 2025, 134 (4): 1259. e1- 1259. e10.

DOI

11
Ke Y , Zhang Y , Tian S , et al. Accuracy of digital implant impressions using a novel structured light scanning system assisted by a planar mirror in the edentulous maxilla: An in vitro study[J]. Clin Oral Implants Res, 2024, 35 (8): 876- 887.

DOI

12
Mourouzis P . Critical methodological factors influencing the accuracy of intraoral scanners in digital dentistry research[J]. Comput Biol Med, 2025, 187, 109780.

DOI

13
Limones A , Cascos R , Molinero-Mourelle P , et al. Impact of the superimposition methods on accuracy analyses in complete-arch digital implant investigation[J]. J Dent, 2024, 147, 105081.

DOI

14
樊铭瑞, 申冰可, 牛文龙, 等. 基于深度学习的多视图立体视觉综述[J]. 软件学报, 2025, 36 (4): 1692- 1714.

15
Li J, Chen Z, Liu F, et al. Obtaining full-arch implant scan with smartphone video and deep learning: An in vitro investigation on trueness and precision [J/OL]. J Prosthodont, 2025 [2025-03-08]. https://pubmed.ncbi.nlm.nih.gov/40055947/.

16
Kernen-Gintaute A , Akulauskas M , Kernen F , et al. Accuracy of digital and conventional implant impressions in edentulous jaws: A clinical comparative study[J]. J Dent, 2025, 153, 105559.

DOI