北京大学学报(医学版) ›› 2023, Vol. 55 ›› Issue (1): 144-148. doi: 10.19723/j.issn.1671-167X.2023.01.022

• 论著 • 上一篇    下一篇

3D打印人工椎体在颈椎脊索瘤全脊椎切除术中的应用

周华,王仁吉,刘忠军,刘晓光,吴奉梁,党礌,韦峰*()   

  1. 北京大学第三医院骨科,骨与关节精准医学工程研究中心,脊柱疾病研究北京市重点实验室,北京 100191
  • 收稿日期:2022-03-25 出版日期:2023-02-18 发布日期:2023-01-31
  • 通讯作者: 韦峰 E-mail:weifeng@bjmu.edu.cn
  • 基金资助:
    北京大学第三医院临床重点项目(BYSY2017001);吴阶平医学基金会临床科研专项基金(320.6750.2022-3-37)

3D-printed vertebral body in anterior spinal reconstruction after total spondylectomy for patients with cervical chordoma

Hua ZHOU,Ren-ji WANG,Zhong-jun LIU,Xiao-guang LIU,Feng-liang WU,Lei DANG,Feng WEI*()   

  1. Department of Orthopaedics, Peking University Third Hospital; Engineering Research Center of Bone and Joint Precision Medicine; Beijing Key Laboratory of Spinal Disease Research, Beijing 100191, China
  • Received:2022-03-25 Online:2023-02-18 Published:2023-01-31
  • Contact: Feng WEI E-mail:weifeng@bjmu.edu.cn
  • Supported by:
    the Key Clinical Projects of Peking University Third Hospital(BYSY2017001);Wu Jieping Medical Foundation Special Clinical Research Project(320.6750.2022-3-37)

摘要:

目的: 通过比较颈椎脊索瘤全脊椎切除术后3D打印人工椎体和钛网下沉率的差异,探讨3D打印人工椎体重建前方椎体能否减少术后内植物的下沉。方法: 回顾性分析2005年3月至2019年9月在北京大学第三医院行全脊椎切除术的24例颈椎脊索瘤患者资料,其中采用3D打印人工椎体重建颈椎前方椎体(3D椎体组)有9例,采用钛网重建(钛网组)有15例。收集患者的年龄、性别、椎体CT值、手术信息(手术节段、手术时间、出血量及围手术期并发症)和术后内植物下沉等随访资料,采用SPSS 22.0进行数据分析。结果: 比较两组患者的性别、年龄、椎体CT值、手术节段、手术时间、出血量及围手术期并发症,差异无统计学意义。术后3个月随访内植物下沉显示,3D椎体组有8例椎体高度下降 < 1 mm (无内植物下沉),1例椎体高度下降>1 mm (轻度内植物下沉);钛网组有5例椎体高度下降 < 1 mm (无内植物下沉),8例椎体高度下沉>1 mm (轻度内植物下沉),2例缺失术后3个月的影像学资料,两组3个月内植物下沉率的比较差异有统计学意义(P < 0.05)。术后12个月随访内植物下沉显示,3D椎体组有8例椎体高度下降 < 1 mm (无内植物下沉),1例椎体高度下沉>3 mm (重度内植物下沉);钛网组有4例椎体高度下降 < 1 mm (无内植物下沉),2例椎体高度下沉>1 mm (轻度内植物下沉),9例椎体高度下沉>3 mm (重度内植物下沉),两组12个月内植物下沉率的比较差异有统计学意义(P < 0.01),且两组的重度内植物下沉率(椎体高度下沉>3 mm) 比较差异有统计学意义(P < 0.05)。术后24个月随访内植物下沉显示,3D椎体组死亡1例,7例椎体高度下降 < 1 mm (无内植物下沉),1例椎体高度下沉>3 mm (重度内植物下沉);钛网组死亡1例,失访1例,1例椎体高度下降 < 1 mm (无内植物下沉),1例椎体高度下沉>1 mm (轻度内植物下沉),11例椎体高度下沉>3 mm (重度内植物下沉),两组24个月内植物下沉率比较差异有统计学意义(P < 0.01),且两组的重度内植物下沉率比较差异有统计学意义(P < 0.01)。结论: 在颈椎脊索瘤的全脊椎切除术中,3D打印人工椎体可提供可靠的即刻和中远期的颈椎稳定性,与钛网重建相比,能够减少术后内植物下沉的发生率。

关键词: 颈椎脊索瘤, 全脊椎切除术, 3D打印人工椎体, 钛网, 内植物下沉

Abstract:

Objective: To investigate whether 3D-printed artificial vertebral body can reduce prosthesis subsidence rate for patients with cervical chordomas, through comparing the rates of prosthesis subsidence between 3D printing artificial vertebral body and titanium mesh for anterior spinal reconstruction after total spondylectomy. Methods: This was a retrospective analysis of patients who underwent surgical treatment for cervical chordoma at our hospital from March 2005 to September 2019. There were nine patients in the group of 3D artificial vertebral body (3D group), and 15 patients in the group of titanium mesh cage (Mesh group). The patients' characteristics and treatment data were extracted from the medical records, including age, gender, CT hounsfield unit of cervical vertebra and surgical information, such as the surgical segments, time and blood loss of surgery, frequency and degree of prosthesis subsidence after surgery. Radiographic observations of prosthesis subsidence during the follow-up, including X-rays, CT, and magnetic resonance imaging were also collected. SPSS 22.0 was used to analysis the data. Results: There was no significant difference between the two groups in gender, age, CT hounsfield unit, surgical segments, time of surgery, blood loss of posterior surgery and total blood loss. Blood loss of anterior surgery was 700 (300, 825) mL in 3D group and 1 500 (750, 2 800) mL in Mesh group (P < 0.05). The prosthesis subsidence during the follow-up, 3 months after surgery, there was significant difference between the two groups in mild prosthesis subsidence (P < 0.05). The vertebral height of the 3D group decreased less than 1 mm in eight cases (no prosthesis subsidence) and more than 1 mm in one case (mild prosthesis subsidence). The vertebral height of the Mesh group decreased less than 1 mm in five cases (no prosthesis subsidence), and more than 1 mm in eight cases (mild prosthesis subsidence). Two patients did not have X-rays in 3 months after surgery. There was a statistically significant difference between the two groups in the prosthesis subsidence rate at the end of 12 months (P < 0.01). The vertebral height of eight cases in the 3D group decreased less than 1 mm (no prosthesis subsidence) and one case more than 3 mm (severe prosthesis subsidence). Four of the 15 cases in the Mesh group decreased less than 1 mm (no prosthesis subsidence), two cases more than 1 mm (mild prosthesis subsidence), and nine cases more than 3 mm (severe prosthesis subsidence). There was a statistically significant difference between the two groups in the prosthesis subsidence rate at the end of 24 months (P < 0.01). The vertebral height of seven cases in the 3D group decreased less than 1 mm (no prosthesis subsidence), one case more than 3 mm (severe prosthesis subsidence), and one case died with tumor. One case in the Mesh group decreased less than 1 mm (no prosthesis subsidence), one case more than 1 mm (mild prosthesis subsidence), 11 case more than 3 mm (severe prosthesis subsidence), one case died with tumor and one lost the follow-up. Moreover, at the end of 12 months and 24 months, there was significant difference between the two groups in severe prosthesis subsidence rate (P < 0.01). Conclusion: 3D-printed artificial vertebral body for anterior spinal reconstruction after total spondylectomy for patients with cervical chordoma can provide reliable spinal stability, and reduce the incidence of prosthesis subsidence after 2-year follow-up.

Key words: Cervical chordoma, Total spondylectomy, 3D-printed artificial vertebral body, Titanium mesh cage, Prosthesis subsidence

中图分类号: 

  • R739.9

图1

颈椎脊索瘤全脊椎切除术后3D打印人工椎体重建"

图2

颈椎脊索瘤全脊椎切除术后钛网重建"

1 Casali PG , Stacchiotti S , Sangalli C , et al. Chordoma[J]. Curr Opin Oncol, 2007, 19 (4): 367- 370.
doi: 10.1097/CCO.0b013e3281214448
2 Dea N , Fisher CG , Reynolds JJ , et al. Current treatment strategy for newly diagnosed chordoma of the mobile spine and sacrum: Results of an international survey[J]. J Neurosurg Spine, 2018, 30 (1): 119- 125.
3 Li Z , Wei F , Liu Z , et al. Risk factors for instrumentation failure after total en bloc spondylectomy of thoracic and lumbar spine tumors using titanium mesh cage for anterior reconstruction[J]. World Neurosurg, 2020, 135, 106- 115.
doi: 10.1016/j.wneu.2019.11.057
4 Wei F , Li Z , Liu Z , et al. Upper cervical spine reconstruction using customized 3D-printed vertebral body in 9 patients with primary tumors involving C2[J]. Ann Transl Med, 2020, 8 (6): 332.
doi: 10.21037/atm.2020.03.32
5 Choy WJ , Mobbs RJ , Wilcox B , et al. Reconstruction of thoracic spine using a personalized 3D-printed vertebral body in adolescent with T9 primary bone tumor[J]. World Neurosurg, 2017, 105, 1013- 1032.
6 Xu N , Wei F , Liu X , et al. Reconstruction of the upper cervical spine using a personalized 3D-printed vertebral body in an adolescent with Ewing sarcoma[J]. Spine (Phila Pa 1976), 2016, 41 (1): 50- 54.
doi: 10.1097/BRS.0000000000001179
7 Wei F , Liu Z , Liu X , et al. An approach to primary tumors of the upper cervical spine with spondylectomy using a combined approach: Our experience with 19 cases[J]. Spine (Phila Pa 1976), 2018, 43 (2): 81- 88.
doi: 10.1097/BRS.0000000000001007
8 Hu B , Wang L , Song Y , et al. A comparison of long-term outcomes of nanohydroxyapatite/polyamide-66 cage and titanium mesh cage in anterior cervical corpectomy and fusion: A clinical follow-up study of least 8 years[J]. Clin Neurol Neurosurg, 2019, 176, 25- 29.
doi: 10.1016/j.clineuro.2018.11.015
9 Jin YZ , Zhao B , Lu XD , et al. Mid- and long-term follow-up efficacy analysis of 3D-printed interbody fusion cages for anterior cervical discectomy and fusion[J]. Orthop Surg, 2021, 13 (7): 1969- 1978.
doi: 10.1111/os.13005
10 Zhou H , Jiang L , Wei F , et al. Prognostic factors in surgical patients with chordomas of the cervical spine: A study of 52 cases from a single institution[J]. Ann Surg Oncol, 2017, 24 (8): 2355- 2362.
doi: 10.1245/s10434-017-5884-5
11 Yoshioka K , Murakami H , Demura S , et al. Risk factors of instrumentation failure after multilevel total en bloc spondylectomy[J]. Spine Surg Relat Res, 2017, 1 (1): 31- 39.
doi: 10.22603/ssrr.1.2016-0005
12 Matsumoto M , Watanabe K , Tsuji T , et al. Late instrumentation failure after total en bloc spondylectomy[J]. J Neurosurg Spine, 2011, 15 (3): 320- 327.
doi: 10.3171/2011.5.SPINE10813
13 Huang S , Ji T , Guo W . Biomechanical comparison of a 3D-printed sacrum prosthesis versus rod-screw systems for reconstruction after total sacrectomy: A finite element analysis[J]. Clin Biomech (Bristol, Avon), 2019, 70, 203- 208.
doi: 10.1016/j.clinbiomech.2019.10.019
14 Kaloostian PE , Gokaslan ZL . Surgical management of primary tumors of the cervical spine: Surgical considerations and avoidance of complications[J]. Neurol Res, 2014, 36 (6): 557- 565.
doi: 10.1179/1743132814Y.0000000367
15 Park SJ , Lee CS , Chang BS , et al. Rod fracture and related factors after total en bloc spondylectomy[J]. Spine J, 2019, 19 (10): 1613- 1619.
doi: 10.1016/j.spinee.2019.04.018
16 Zhang T , Wei Q , Zhou H , et al. Three-dimensional-printed individualized porous implants: A new "implant-bone" interface fusion concept for large bone defect treatment[J]. Bioact Mater, 2021, 6 (11): 3659- 3670.
doi: 10.1016/j.bioactmat.2021.03.030
17 Xiu P , Jia Z , Lv J , et al. Tailored surface treatment of 3D printed porous Ti6Al4V by microarc oxidation for enhanced osseointegration via optimized bone in-growth patterns and interlocked bone/implant interface[J]. ACS Appl Mater Interfaces, 2016, 8 (28): 17964- 17975.
doi: 10.1021/acsami.6b05893
18 Girolami M , Boriani S , Bandiera S , et al. Biomimetic 3D-printed custom-made prosthesis for anterior column reconstruction in the thoracolumbar spine, a tailored option following en bloc resection for spinal tumors: Preliminary results on a case-series of 13 patients[J]. Eur Spine J, 2018, 27 (12): 3073- 3083.
doi: 10.1007/s00586-018-5708-8
19 Yang J , Cai H , Lv J , et al. In vivo study of a self-stabilizing artificial vertebral body fabricated by electron beam melting[J]. Spine (Phila Pa 1976), 2014, 39 (8): 486- 492.
doi: 10.1097/BRS.0000000000000211
[1] 马忠泰, 施学东, 栗怀广. 全脊椎切除术治疗脊柱转移癌[J]. 北京大学学报(医学版), 2002, 34(6): 645-652.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
[1] 王学庆, 万有, 于英心, 韩济生. 关节炎大鼠背根神经节细胞的膜电生理学特征[J]. 北京大学学报(医学版), 2001, 33(1): 50 -53 .
[2] 唐志慧, 曾祥龙. 恒牙早期正常骨面型青少年上气道形态和舌骨位置的X线头影测量研究[J]. 北京大学学报(医学版), 2002, 34(2): 140 -143 .
[3] 关志忱, 魏本林, 孟作为. 远程无线排尿日记开发及20例年轻人客观排尿情况报告[J]. 北京大学学报(医学版), 2010, 42(4): 476 -479 .
[4] 张春丽, 王荣福, 李太华, 付占立. 新型有机锗倍半氧化物的抗肿瘤活性及其在荷瘤裸鼠的体内分布[J]. 北京大学学报(医学版), 2008, 40(2): 208 -210 .
[5] 胡维亨, 任军. 人乙型肝炎病毒DNA阳性血清对人骨髓间充质干细胞向肝细胞分化的影响[J]. 北京大学学报(医学版), 2008, 40(5): 459 -464 .
[6] 王军, 肖水芳, 秦永, 王全桂, 陈丽. 以面神经麻痹为首诊表现的Wegener肉芽肿病一例[J]. 北京大学学报(医学版), 2007, 39(4): 434 -436 .
[7] 柳晓辉, 那加, 刘玲玲, 罗斌. 头颈部血管肉瘤3例[J]. 北京大学学报(医学版), 2001, 33(3): 288 -289 .
[8] 张震康. 口腔医学科学研究的重要进展和方向[J]. 北京大学学报(医学版), 2002, 34(2): 97 -98 .
[9] 梁成, 王兴, 伊彪, 李自力, 王晓霞. 骨性颞下颌关节强直伴小颌畸形及阻塞性睡眠呼吸暂停综合征的牵引成骨治疗[J]. 北京大学学报(医学版), 2002, 34(2): 112 -116 .
[10] 张勇, 栾庆先. 牙周维护治疗在保持牙周长期疗效中的作用[J]. 北京大学学报(医学版), 2011, 43(1): 29 -33 .