一种数字化全口义齿的临床疗效评价

魏菱1,邹东2,陈虎1,潘韶霞1,(),孙玉春1,(),周永胜1   

  1. 1.北京大学口腔医学院·口腔医院,修复科 国家口腔疾病临床医学研究中心 口腔数字化医疗技术和材料国家工程实验室 口腔数字医学北京市重点实验室,北京 100081
    2.中日友好医院口腔医学中心,北京 100029
  • 收稿日期:2019-09-19 出版日期:2020-08-18 发布日期:2020-08-06
  • 通讯作者: 潘韶霞,孙玉春 E-mail:panshaoxia@vip.163.com;polarshining@163.com
  • 基金资助:
    首都卫生发展科研专项(2016-1-4101);北京大学口腔医院临床新技术新疗法项目(PKUSSNCT-15A09)

Evaluation of clinical efficacy of a kind of digital complete denture

Ling WEI1,Dong ZOU2,Hu CHEN1,Shao-xia PAN1,(),Yu-chun SUN1,(),Yong-sheng ZHOU1   

  1. 1. Department of Prosthodontics, Peking University School and Hospital of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Laboratory for Digital and Material Technology of Stomatology & Beijing Key Laboratory of Digital Stomatology, Beijing 100081, China
    2. Centre of Stomatology, China-Japan Friendship Hospital, Beijing 100029, China
  • Received:2019-09-19 Online:2020-08-18 Published:2020-08-06
  • Contact: Shao-xia PAN,Yu-chun SUN E-mail:panshaoxia@vip.163.com;polarshining@163.com
  • Supported by:
    Capital Health Research and Development of Special(2016-1-4101);Program for New Clinical Techniques and Therapies of Peking University School and Hospital of Stomatology(PKUSSNCT-15A09)

摘要:

目的: 比较计算机辅助设计与制造(computer aided design/computer aided manufacturing,CAD/CAM)技术制作的数字化全口义齿与传统方法制作的全口义齿的临床疗效。方法: 纳入20例无牙颌患者,通过前瞻性、单盲、自身对照临床实验,每人分别采用功能易适性数字化义齿(functional suitable denture, FSD)制作系统及传统方法制作全口义齿。义齿完成后,首先戴用传统全口义齿, 3个月后交换戴用数字化全口义齿。患者于修复前、修复后第2周、第1个月、第2个月、第3个月填写无牙颌口腔健康影响量表-20(oral health impact profile edentulous,OHIP-20E)和义齿满意度视觉模拟评分量表(visual analogue scale,VAS)。于两副义齿戴用后的第3个月,分别进行咀嚼效率测试。结果: 20例患者在戴牙后第3个月,FSD全口义齿VAS评分在总体满意度、易清洁性、发音、美观、稳定、口腔状况这6个方面均值均高于传统全口义齿,而在舒适、咀嚼能力、嚼碎功能这3个方面VAS评分的均值与传统全口义齿基本持平,但差异无统计学意义(P>0.05);FSD全口义齿OHIP-20E评分在各个方面的均值均高于传统全口义齿,其中功能受限、心理不适、总评分这3个方面评分差异有统计学意义(P<0.05);咀嚼效率测量结果为,FSD全口义齿平均咀嚼效率(1.20±0.54)高于传统全口义齿的平均咀嚼效率(1.16±0.53), 差异无统计学意义(P=0.691)。结论: 本研究评价的数字化全口义齿的疗效与传统全口义齿相同,在患者满意度及口腔相关生活质量评价方面都展现出相同的治疗效果。

关键词: 计算机辅助设计与制造, 功能易适性数字化义齿系统, 全口义齿, 临床疗效

Abstract:

Objective: To compare the clinical efficacy of a kind of complete dentures fabricated using computer aided design/computer aided manufacturing (CAD/CAM) workflow with that of the complete denture fabricated using conventional workflow. Methods: Twenty edentulous patients were included in this prospective, single-blind, self-controlled clinical trial. Two pairs of complete dentures were fabricated for each participant: one using the functional suitable denture (FSD) system with CAD/CAM, and the other using conventional fabrication workflow. In the final delivery appointment, the conventional dentures worn by the participant for 3 months at first. Then the participant swapped to wear digital denture for another 3 months. The patients’ satisfaction and oral health related quality of life (OHRQoL) were measured using the visual analogue scale (VAS) and oral health impact profile (OHIP-20E) at baseline, 2 weeks, 1 month, 2 months, and 3 months following denture delivery. In addition, the masticatory efficiency of each pair of dentures was measured after three months adaptation period for each type of dentures. Results: The VAS scores of the twenty patients in FSD denture group on general satisfaction, ease of cleaning, ability to speak, esthetics, stability and oral health status on these six domains were higher than that of conventional denture group 3 months after delivery. While the VAS cores in FSD denture group on comfort, ability to chew and ability to chew up were the same as those of conventional denture group. But there was no significant difference (P>0.05). The VAS score on each domain in FSD denture group was higher than that of conventional denture group 3 months after delivery, three domains of which reached statistical significance (P<0.05), involving functional limitation, psychological discomfort and total. The masticatory performance of the FSD denture (1.20±0.54) was slightly higher than that for the conventional denture (1.16±0.53), but the difference did not reach statistical significance (P=0.691). Conclusion: The clinical efficacy of the FSD complete denture is comparable to that of the conventional complete denture. As for patient satisfaction and oral health related quality of life, FSD dentures received comparable scores as conventional complete dentures did.

Key words: Computer aided design/computer aided manufacturing, Functional suitable denture system, Complete denture, Clinical efficacy

中图分类号: 

  • R783.2

图1

FSD全口义齿与传统全口义齿临床疗效比较的实验设计流程图"

图2

基线与戴用FSD全口义齿后2周、1个月、2个月、3个月各方面VAS得分对比图"

图3

基线与戴用传统全口义齿后2周、1个月、2个月、3个月各方面VAS得分对比图"

图4

各观察时点患者戴用FSD和传统全口义齿总体义齿满意度VAS评分"

图5

OHIP-20E基线与戴FSD全口义齿后2周、1个月、2个月及3个月各分项得分对比图"

图6

OHIP-20E基线与戴传统全口义齿后2周、1个月、2个月及3个月各分项得分对比图"

图7

20例患者分别戴用两种全口义齿各观察时间节点OHIP-20E总评分"

[1] Feine JS, Carlsson GE, Awad MA, et al. The McGill consensus statement on overdentures[J]. Int J Prosthodont, 2002,15(4):413-414.
[2] Heydecke G, Locker D, Awad MA, et al. Oral and general health-related quality of life with conventional and implant dentures[J]. Community Dent Oral Epidemiol, 2003,31(3):161-168.
[3] Awad MA, Lund JP, Dufresne E, et al. Comparing the efficacy of mandibular implant-retained overdentures and conventional dentures among middle-aged edentulous patients: satisfaction and functional assessment[J]. Int J Prosthodont, 2003,16(2):117-122.
[4] Emami E, Heydecke G, Rompre PH, et al. Impact of implant support for mandibular dentures on satisfaction, oral and general health-related quality of life: a meta-analysis of randomized-controlled trials[J]. Clin Oral Imp, 2009,20(6):533-544.
[5] Andreiotelli M, Att W, Strub JR. Prosthodontic complications with implant overdentures: a systematic literature review[J]. Int J Prosthodont, 2010,23(3):195-203.
pmid: 20552083
[6] Bozini T, Petridis H, Garefis K, et al. A meta-analysis of prosthodontic complication rates of implant-supported fixed dental prostheses in edentulous patients after an observation period of at least 5 years[J]. Int J Oral Maxillofac Implants, 2011,26(2):304-318.
pmid: 21483883
[7] Bidra AS, Farrell K. Prospective cohort pilot study of 2-visit CAD/CAM monolithic complete dentures and implant-retained overdentures: clinical and patient-centered outcomes[J]. J Prosthet Dent, 2016,115(5):578-586.
[8] Douglass CW, Shih A, Ostry L. Will there be a need for complete dentures in the United States in 2020[J]. J Prosthet Dent, 2002,87(1):5-8.
pmid: 11807476
[9] Miyazaki T, Hotta Y, Kunii J, et al. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience[J]. Dent Mater J, 2009,28(1):44-56.
pmid: 19280967
[10] Wimmer T, Gallus K, Eichberger M, et al. Complete denture fabrication supported by CAD/CAM[J]. J Prosthet Dent, 2016,115(5):541-546.
pmid: 26774323
[11] Matsuda T, Goto T, Yagi K, Kashiwabara T, et al. Part-digitizing system of impression and interocclusal record for complete denture fabrication[J]. J Prosthodont, 2016,25(6):503-509.
[12] Goodacre CJ, Garbacea A, Naylor WP, et al. CAD/CAM fabricated complete dentures: concepts and clinical methods of obtaining required morphological data[J]. Prosthet Dent, 2012,107(1):34-46.
[13] Inokoshi M, Kanazawa M, Minakuchi S. Evaluation of a complete denture trial method applying rapid prototyping[J]. Dent Mater J, 2012,31(1):40-46.
doi: 10.4012/dmj.2011-113
[14] Zhang YD, Jiang JG, Liang T, et al. Kinematics modeling and experimentation of the multi-manipulator tooth-arrangement robot for full denture manufacturing[J]. J Med Syst, 2011,35(6):1421-1429.
pmid: 20703775
[15] Kanazawa M, Inokoshi M, Minakuchi S, et al. Trial of a CAD/CAM system for fabricating complete dentures[J]. Dent Mater J, 2011,30(1):93-96.
doi: 10.4012/dmj.2010-112
[16] Bidra AS, Taylor TD. Computer-aided technology for fabricating complete dentures: systematic review of historical background, current status, and future perspectives[J]. J Prosthet Dent, 2013,109(6):361-366.
doi: 10.1016/S0022-3913(13)60318-2 pmid: 23763779
[17] 孙玉春, 吕培军, 王勇. 参数化定位全口义齿人工牙三维图形数据库的研究[J]. 口腔颌面修复学杂志, 2007,8(4):257-260.
[18] 王勇, 吕培军, 赵占芳. 用于全口义齿的人工牙三维坐标系的建立[J]. 实用口腔医学杂志, 2005,21(5):653-655.
[19] 孙玉春, 吕培军, 王勇. 计算机辅助设计与快速成形技术辅助制作全口义齿的探讨[J]. 中华口腔医学杂志, 2007,42(6):324-329.
[20] 孙玉春, 吕培军, 王勇. 用于全口义齿计算机辅助设计的虚拟半可调牙合架[J]. 北京大学学报(医学版), 2008,40(1):92-96.
[21] Sun YC, Lv PJ, Wang Y. Study on CAD & RP for removable complete denture[J]. Comput Meth Prog Bio, 2009,93(93):266-272.
[22] Chen H, Wang H, Lv PJ. Quantitative evaluation of tissue surface adaption of CAD-designed and 3D printed wax pattern of maxillary complete denture[J]. Biomed Res Int, 2015,2015:453968.
pmid: 26583108
[23] Chen H, Yang X, Chen L, et al. Application of FDM three-dimensional printing technology in the digital manufacture of custom dentulous mandible trays[J]. Sci Rep, 2016,6:19207.
doi: 10.1038/srep19207 pmid: 26763620
[24] Yuan FS, Lv PJ, Wang PF, et al. Custom fabrication of try-inwax complete denture[J]. Rapid Prototyp J, 2016,22(3):539-543.
[25] Critchlow SB, Ellis JS. Prognostic indications for conventional complete denture therapy: a review of the literature[J]. J Dent, 2010,38(1):2-9.
doi: 10.1016/j.jdent.2009.08.004 pmid: 19695302
[26] 冯海兰, 徐军. 口腔修复学[M]. 2版. 北京: 北京大学医学出版社, 2015: 294.
[27] Reeve PE. The role of personality in the management of complete denture patients[J]. Br Dent J, 1984,156(10):356-362.
doi: 10.1038/sj.bdj.4805366 pmid: 6587876
[28] Slade GD, Spencer AJ. Development and evaluation of the oral health impact profile[J]. Community Dent Health, 1994,11(1):3-11.
pmid: 8193981
[29] Awad MA, Lund JP, Shapiro SH, et al. Oral health status and treatment satisfaction with mandibular implant overdentures and conventional dentures: a randomized clinical trial in a senior population[J]. Int J Prosthodont, 2003,16(4):390-396.
pmid: 12956494
[30] Bolender CL. The Cornell Medical Index as prognostic aid for complete denture patients[J]. J Prosthet Dent, 1969,22(1):20-29.
doi: 10.1016/0022-3913(69)90268-6 pmid: 5254404
[31] Vervoorn JM, Duinkerke AS, Luteijn F, et al. Assessment of denture satisfaction[J]. Community Dent Oral Epidemiol, 1988,16(6):364-367.
doi: 10.1111/j.1600-0528.1988.tb00583.x pmid: 3060310
[32] Bear ML. The use of the psychological measures in predicting patient satisfaction with complete dentures[J]. Int J Prosthodont, 1992,5(3):221-226.
pmid: 1524644
[33] Kalk W, Baat C. Patients’ complaints and satisfaction 5 years after complete denture treatment[J]. Community Dent Oral Epidemiol, 1990,18(1):27-31.
pmid: 2404654
[1] 周团锋,杨雪,王睿捷,程明轩,张华,韦金奇. 数字化制作简易口内哥特式弓在全口义齿修复正中关系确定中的应用[J]. 北京大学学报(医学版), 2023, 55(1): 101-107.
[2] 蒋艳芳,王健,王永健,刘佳,裴殷,刘晓鹏,敖英芳,马勇. 前交叉韧带翻修重建术后中长期临床疗效及影响因素[J]. 北京大学学报(医学版), 2021, 53(5): 857-863.
[3] 孙玉春,王勇,邓珂慧,陈虎,李伟伟,赵一姣,潘韶霞,叶红强,周永胜. 功能易适数字化全口义齿的自主创新研发[J]. 北京大学学报(医学版), 2020, 52(2): 390-394.
[4] 刘冰川,杨钟玮,周方,姬洪全,张志山,郭琰,田耘. 肱骨近端骨折微创锁定钢板改良内固定的疗效分析[J]. 北京大学学报(医学版), 2019, 51(2): 277-282.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
[1] 李云芳, 张幼怡, 侯嵘, 董尔丹, 韩启德. 质粒转染对HEK293和DDT1-MF2细胞天然β2-肾上腺素受体表达的影响[J]. 北京大学学报(医学版), 2001, 33(5): 457 -461 .
[2] 柯杨. 乳头状瘤病毒与人类肿瘤[J]. 北京大学学报(医学版), 2002, 34(5): 599 -603 .
[3] 李智岗, 黄景香, 李顺宗, 赵俊京, 时高峰, 梁国庆, 王红光, 韩捧银, 王琦, 谷铁树. 肝转移瘤的血供[J]. 北京大学学报(医学版), 2008, 40(2): 146 -150 .
[4] 冯现竹, 侯平, 朱厉, 于磊, 张宏. 转铁蛋白受体基因多态性与IgA肾病易感性及临床病理表型的相关性[J]. 北京大学学报(医学版), 2008, 40(4): 369 -373 .
[5] 王倩, 张翼, 陆敏, 管又飞, 朱毅, 王悦. 高盐诱导的高血压大鼠模型肾组织可溶性表氧化物酶高表达及其作用初步探讨[J]. 北京大学学报(医学版), 2010, 42(2): 126 -130 .
[6] 管宏, 赵慧云, 沈磊, 李五岭, 王建华, 王春荣, 徐福. 联合应用重组TPO和G-CSF对骨髓抑制性小鼠外周血小板及白细胞恢复的影响[J]. 北京大学学报(医学版), 2001, 33(2): 181 -182 .
[7] 赵建新, 周良, 万远廉. 经十二指肠逆行放置支架治疗恶性幽门梗阻2例[J]. 北京大学学报(医学版), 2002, 34(6): 737 -738 .
[8] 牟向东, 王广发, 阙呈立, 李桂莲. H3N2型人流行性感冒合并金黄色葡萄球菌败血症及金黄色葡萄球菌肺炎1例[J]. 北京大学学报(医学版), 2007, 39(6): 663 -665 .
[9] 王晓霞, 王兴, 伊彪, 李自力, 梁成. 上颌骨牵引成骨术对腭裂患者腭咽部结构的影响[J]. 北京大学学报(医学版), 2005, 37(6): 648 -651 .
[10] 黄尚志, 高翼之. 中国医学遗传学发展史[J]. 北京大学学报(医学版), 2006, 38(1): 9 -13 .